32?channel programmable digital I/O module
Key Facts
  • 32 digital input and 32 digital output channels
  • 40 MHz sample rate, depending on levels and number of channels
  • FPGA-based flexibility and realtime task execution independent of operating system used
  • 12.5 ns resolution
  • 8 groups of 4 channels each: –6 V to +10 V
  • Tristate control for every output channel in dynamic mode
  • Programmable DIO level (high and low) per group with 14 bit resolution
  • Two programmable input threshold levels per group for hysteresis or level monitoring
  • High output current with 150 mA per channel
  • Synchronization/triggering (bidirectional) via PXI trigger bus or XTI (TTL)
  • External clock input via EXT_CLK input pin (TTL)
  • Selftest software
  • Soft front panels for immediate use
  • LabWindows/CVI device driver support
  • Test software library R&S®GTSL in DLL format

Brief Description The R&S®TS-PIO4 digital functional test module contains flexibly programmable 32-bit digital inputs and 32-bit digital outputs that are able to acquire and stimulate static or dynamic digital patterns. The R&S®TS-PIO4 digital functional test module is a compact PCI/PXI module that takes up only one slot on the R&S®CompactTSVP test system versatile platform. The characteristics of the digital lines can be configured with 14 bit resolution levels per port. In addition, the levels of the output ports can be programmed to meet application requirements. The input ports have a programmable threshold and hysteresis to match the specifications of common digital logic families.
Specifications
Use on R&S®CompactTSVP platform
R&S®CompactTSVP1 slot required
Interfaces
Control busCompactPCI/PXI
DUT connector (front)DIN 41612, 96 pins
Rear I/O connectorCompactPCI connector, 110 pins
Module features
Data input channels
Channelssingle-ended channels32, on 8 ports of 4 bit
Input modes 1)hysteresis, comparator
Input levels 2)clamping if outside range
range 0–7.5 V to +10.0 V
range 1–5.0 V to +12.0 V
range 2–3.3 V to +12.0 V
range 30.0 V to +12.0 V
Input thresholds 2)two programmable thresholds, 14 bit resolution
range 0–6.0 V to +5.1 V
range 1–3.5 V to +7.1 V
range 2–1.8 V to +7.1 V
range 3+1.5 V to +7.1 V
Input resistance1 M?
Signal acquisition timing
Sample rate0.01 Hz to 40 MHz with 12.5 ns resolution, depending on levels and number of channels
Trigger delay50 ns to 50 s with 12.5 ns resolution
Data buffer depth/widthprogrammable2 Msample at 32 bit (IN 1 to IN 32)
Protection
Overvoltage protection±24 V
Data output channels
Channelssingle-ended channels32, on 8 ports of 4 bit, the number of channels that can be simultaneously used depends on the level range, frequency and load 2)
Output modesanalog, tristate
Analog output mode 2)output voltage
range 0VL = –6.0 V to –1.5 V
VH = –6.0 V to +8.0 V
range 1VL = –3.5 V to +1.0 V
VH = –3.5 V to +10.0 V
range 2VL = –1.8 V to +2.7 V
VH = –1.8 V to +10.0 V
range 3VL = +1.5 V to +6.0 V
VH = +1.5 V to +10.0 V
output currentmax. 150 mA per channel, 10 mA to 350 mA per port 1)
output resolution14 bit
Tristate control output modeoutput voltagenone
Output resistancetyp. 30 ?
Realtime stimulation 1)
Sample rate0.01 Hz to 40 MHz at 12.5 ns resolution, depending on levels and number of channels
Trigger delay50 ns to 50 s at 12.5 ns resolution
Data buffer depth/widthprogrammable2 Msample at 32 bits (OUT 1 to OUT 32)
Implementation of bidirectional data busestristate controlprogrammable, per sample, portwise connection of data input with data output channels via onboard analog switches
Protectionshort-circuitreverse voltage up to ±24 V at 150 mA
Relay channel
Ground-relay channelground relay to disconnect chassis ground for in?circuit tests (ICT)SPST, to ground
Maximum switching voltageDC/AC30 V/125 V (RMS)
Maximum switching currentDC/AC2 A/0.5 A (RMS)
Maximum switching powerDC/AC60 W/62.5 VA
Pattern comparator
Comparison32-bit data input channel pattern with 32-bit reference pattern, 32-bit comparison enable mask
Applications
Frequency measurement 3)maximum input frequency10 MHz at 50 % duty cycle
minimum pulse width50 ns
frequency resolution25 ns
measurement time200 ns to 50 s gate time or up to 232 – 1 comparator matches
triggersoftware trigger
Event counting 3)minimum pattern duration50 ns
gate time200 ns to 50 s
event countsup to 232 – 1
Trigger generatoroutput to PXI triggerXTO input to trigger units
Synchronization
Trigger units2, fully independent hardware trigger logic
Applications? programmable trigger generator ? generation of realtime stimulation clock ? generation of realtime acquisition clock ? frequency measurement
Synchronization inputs? 1, local TTL trigger (XTI) ? 8, PXI trigger bus ? 1, pattern comparator
Trigger unit characteristics 4)reference pattern10 bit, 3 states: high, low, don’t care
slopepositive/negative
delay25 ns to 50 s
output signals? trigger received signal (25 ns pulse) ? trigger active signal (start of trigger until end of burst) ? sample pulse (25 ns pulse for each sample)
Synchronization outputs? 1, local TTL trigger (XTO) ? 8, PXI trigger bus
Signals? output, trigger unit 1 (IT1) ? output, trigger unit 2 (IT2) ? output, pattern comparator ? input signal to trigger unit (XTI)
External clock
Input voltage5 V TTL
Frequency range20 MHz to 40 MHz
General data
Power consumption40 W, mainly powered from 3.3 V/typ. 2.5 A and 5 V/typ. 4.5 A 5)
Environmental conditions
Temperatureoperating temperature range+5 °C to +40 °C
storage temperature range–10 °C to +60 °C
Damp heat+40 °C, 80 % rel. humidity, steady state, in line with EN 60068-2-30
Mechanical resistance
Vibrationsinusoidal5 Hz to 55 Hz, 0.15 mm const. amplitude, 55 Hz to 150 Hz, 0.5 g const., in line with EN 60068-2-6
random10 Hz to 300 Hz, acceleration 1.2 g (RMS), in line with EN 60068-2-64
Shock40 g shock spectrum, in line with MIL-STD-810E, method 516.4
Product conformity
Electromagnetic compatibilityEU: in line with EMC Directive 2004/108/ECapplied harmonized standards: EN 61326-1 (industrial environment), EN 61326-2-1, EN 55011 (class A), EN 61000-3-2, EN 61000-3-3
Electrical safetyEU: in line with Low Voltage Directive 2006/95/ECapplied harmonized standard: EN 61010-1
DimensionsW ? H ? D20 mm ? 174 mm ? 316 mm (0.8 in ? 6.8 in ? 12.4 in)
Weight0.75 kg (1.65 lb)
1) Programmable per group. 2) For an estimation of the limits in ranges 0 to 3, see R&S®TS-PIO4 user manual, chapter 5.3.7 Power consumption. If the operating mode to be used exceeds 80 % of the maximum allowed power consumption, monitoring of the I/O driver temperature is recommended. 3) Uses one trigger unit. 4) Programmable per trigger unit. 5) Output voltage range 2 with VL = –1.8 V to +2.7 V, VH = –1.8 V to +10.0 V, 16 active outputs, frequency 20 MHz, sample rate 40 Mbps, no load (see user manual, chapter 5.3.7 “Power consumption”).